General features and scope
|
|
|
Range of water contents
Kind of samples
Determination of the bromine number
Input for separate reference electrode
Number of ready-for-use methods
Storage capacity of user methods
Automation with 760 KF Sample Changer
Automation with 774 Oven Sample Processor
|
100 ppm...100%
solid, liquid and gaseous
yes
yes
18
up to 100
yes
yes
|
Determination
|
|
Automatic conditioning
Range of start volume
Drift stop as endpoint criterion
Stop delay as endpoint criterion
Drift determination
Autostart after sample addition
|
yes
0...99,99 mL
yes
yes
yes
yes
|
| Result calculation and output |
|
Number of results per method
Statistics (up to 20 determinations)
Curve printout
GLP-compliant reports |
9
mean value, absolute and relative standard deviation
yes
yes
|
| Interfaces |
|
2 RS 232 connections for balance, printer or PC
Remote connection
stirrer connection
Connection for additional dosing elements for auxiliary solutions (776,
765) - back titration
|
yes
yes
yes
yes
|
| Hardware |
|
Backlit graphics LCD screen
for the instrument dialog, display of real-time curves,
parameters and results
Separate keypad
2 titration vessels
Double platinum electrode
Measuring input for polarized electrodes |
yes
yes
20
90 mL, 50...150 mL
yes
yes
|
| |
|
| Dialogue Languages |
English, German, French, Spanish, Italian,
Portuguese, Swedish
|
| Dimensions |
|
Width
Height (incl. Exchange Unit)
Depth
Weight (incl. accessories) |
150 mm
450 mm
275 mm
6.6 kg
|