787 KF Titrino
The reasonably priced KF all-round titrator for routine
analysis
|
|
Once installed, the 787 KF Titrino is immediately ready
for use as it contains the four most important titration
methods. The individual methods are arranged such that
the following determinations can be performed using the
standard parameter:
- Titer determination with water or water standard
- Titer determination with sodium tartrate
- Blank determination
- Karl Fischer titration - KFT
Approximately 90 % of all applications can be carried
out using these standard parameters. Of course the parameters
can be adapted to the sample and the KF reagent at any
time, e.g. when n auxiliary solvent is used to enhance
sample dissolution.
|
Direct water content determinations in
- Liquids
- Solids
- Gases
- Viscous materials
Water content determination by the Karl Fischer method
offers you
- rapid results (typically three minutes) through unique
drift stop control,
- accurate results as reaction is specific,
- «odourless» results as work is performed
in a closed system.
|
Routine operation
The 787 KF Titrino applies the Titrino concept to water
content determination. It combines extremely
simple routine operation with outstanding versatility.
Multitasking allows additional functions parallel to
titration. Entries via the separate keypad or via the built-in
RS 232C interface are possible even during titration. For routine analysis, the keypad can be detached and
the instrument operated using three keys on the front
panel. An easy-to-operate PC database for data backup and date
reprocessing is included: Metrodata VESUV Light. 787 KF Titrino keypad
|
Progress in KF methodology
The 787 KF Titrator combines extremely easy routine
operation with flexibility, thus allowing you to adapt
the instrument configuration to your special requirements:
- Apply bi-voltametric or bi-amperometric endpoint
indication with freely selectable polarizing current
or voltage.
- Choose between two stop criteria, namely the conventional "delay
time" or the novel "drift stop", which,
combined with the high resolution of the buret volume
(10 000 increments), yields fast yet highly precise
results.
- Define a start volume to further speed up the determinations.
- Take advantage of the "live keyboard" to
change parameters while titration is in progress.
- Work with or without solution conditioning between
determinations, with or without an automatic request
for sample size and sample identification.
- In the "conditioning" mode, get the current
drift value without delay at the push of a button.
- Have drift taken into account for result calculation.
|
If there is moisture in
it, Metrohm titrators will find it
|
Get more information
If you wish to get more information you may either submit
our reply
form or send an e-mail to info@metrohm.com.
|
| For each Titrando and Titrino you can order a file containing practical
validation documentation. This is part of the comprehensive validation
services offered by your local Metrohm distributor.
|
Ordering information |
|
2.787.0020
|
787 KF Titrino
Compact KF Titrator for fast and precise water determinations. Titration
stop either by means of delay time or by drift criterion. Software-controlled
polarizer for bi-voltametric or bi-amperometric indication. Easy operation
thanks to dialogue in English, French, Spanish or German. Equipped with
RS 232C serial interface and remote control (I/O lines). Including accessories.
|
|
|
Options |
| 6.5613.000 |
ECO semi-micro titration equipment |
| 6.0234.100 |
Combined "Long Life" micro pH electrode for eco
semi-micro titration equipment |
| 6.0338.100 |
Double Pt electrode with plug-in head, for eco Karl Fischer
titrations |
| 6.2104.020 |
Electrode cable with F plug, length 1 m |
| 6.2104.030 |
Electrode cable with F plug, length 2 m |
Additional dispensing units
|
2.776.0010
|
776 Dosimat as auxiliary Dosimat |
| 2.765.0010 |
765 Dosimat as auxiliary Dosimat |
| 6.2139.000 |
Connecting cable Titrino - 776 or 765 Dosimat |
Remote Box
|
6.2148.000
|
Remote Box for connecting PC keyboard and barcode reader |
Stirrer, titration stands
|
2.728.0040
|
728 Magnetic Stirrer with stand, electrode holder and stirring bar |
2.727.0010
|
727 Titration Stand with rinsing pump and rinsing head;
requires 2.722.0010 |
| 2.802.0010 |
802 Rod Stirrer for 2.727.0010, with controller and stirrer propeller |
| 2.727.0100 |
727 Titration Stand with controller and built-in magnetic stirrer, rinsing
pump and rinsing head |
| 2.703.0010 |
703 Titration Stand
Highly integrated, manually operated titration stand with stirrer and pump
for addition of solvents and aspiration of spent solution when performing
endpoint titrations (SET) with conditioning between determinations |
Exchange Units with glass cylinder and flat PCTFE/PTFE stopcock
|
6.3026.110
|
1 mL Exchange Unit with one tip each for titration and dosing |
|
6.3026.150
|
5 mL Exchange Unit with one tip each for titration and dosing |
| 6.3026.210 |
10 mL Exchange Unit with one tip each for titration and dosing |
| 6.3026.220 |
20 mL Exchange Unit with one tip each for titration and dosing |
| 6.3026.250 |
50 mL Exchange Unit with one tip each for titration and dosing |
| |
The bottle attachment of the above Exchange Units allow direct reagent
dispensing from bottles with GL45 glass threads. For all other types, the
following adapters are available: |
| 6.1602.110 |
Siphon for Merck reagent bottles with S40 thread |
| 6.1618.000 |
Adapter 32 mm/GL45 |
| 6.1618.010 |
Adapter 28 mm/GL45 |
| |
|
| 6.1542.010 |
Flat ceramic stopcock for 6.3026.XX0 Exchange Units |
| 6.1542.020 |
Replacement stopcock, PCTFE/PTFE, for 6.3026.XX0 Exchange
Unit |
Printer
|
2.140.0200
|
Custom DP40 impact printer, paper width 58 mm, serial RS 232C
interface (DB9) |
| 6.9001.040 |
Paper roll for Custom DP40 |
| 6.9002.050 |
Ink ribbon for Custom DP40 |
Metrodata VESUV© 3.0
|
| 6.6008.200 |
Metrodata VESUV 3.0, including Hardware Dongle, for n instruments |
| PC program (Windows 95, 98, NT, 2000
or XP) for the acquisition of titration data via an RS 232C interface, method
backup, printing of reports, curve display and reprocessing of data. Data
base with filter, search and query functions, data export to Excel, Lotus,
LIMS
Dialog in English or German. |
Connecting cables for printers, balances and PCs
|
| 6.2134.110 |
Connecting cable for Seiko DPU 414 or Custom DP40 printers |
| 6.2134.050 |
Connecting cable for Epson printers FX, LX, LQ |
| 6.2125.040* |
Cable for Epson EX 800/LQ 850 printers (DIN plug) |
| 2.145.0300 |
Serial-parallel converter SP 1000 for connecting the 798 MPT
Titrino to a printer having a parallel interface (Centronics); requires
6.2125.020 connecting cable and 6.2125.010 adapter |
| 6.2125.030* |
Stackable plug for the simultaneous connection of printer
and balance |
| 6.2134.060 |
Cable for Sartorius balances MP8 and MC1 |
| 6.2134.110 |
Cable for Mettler balances PG, AB..S |
| 6.2125.020* |
Cable for Mettler balances AE (011/012) as well as for AND
balances (for Mettler balances AM, AT, PM use Mettler cable ME33995 and
for Mettler balances AB, AG Mettler interface LC-RS25) |
| 6.2125.080* |
Cable for Precisa balances |
| 6.2125.060 |
Cable for PC with 25-pin connector (DB25) |
| 6.2125.110 |
Cable for PC with 9-pin connector (DB 9) |
| |
|
|
|
|
 |
|
|
|
|
|