| The procedure
could not be simpler. Depending on the type of Metrohm Titrino,
you have two different possibilities:
- DMP-Titrino 785
or GPD-Titrino 751
(instrument program version 5.751.0020 or higher)
Using the method memory card, you load all the methods
(complete parameter sets including calculations) into your Titrino.
You then equip the Titrino with the corresponding Exchange Unit
and electrodes, load the desired titration method into the working
memory, prepare the first sample and start the determination by
pressing a key. As soon as the titration is concluded and the
sample weight entered, the Titrino outputs the titration curve
and a complete result report on the connected printer.
- DMS-Titrino 716,
GP-Titrino 736 or GPD-Titrino 751 (instrument program version
earlier than 5.751.0020)
Using a PC and the demo version of Metrodata VESUV 3
contained in the package, you load the methods into your Titrino.
You then continue using the extremely simple procedure described
above.
To use the methods with
a Titrando, import
the methods. The translation program of the software will automatically
convert the methods. |
|
With
Metrohm you can choose between five different surfactant electrodes
depending on your particular application: from left to right
the
6.0507.010 NIO electrode,
the 6.0504.150 "High Sense" surfactant electrode,
the 6.0507.120 "Ionic Surfactant" electrode,
the 6.0507.130 Surfactrode Resistant and
the 6.0507.140 Surfactrode Refill .
|
|