Antimony(III) in zinc plant electrolyte with chloranilic acid as complexing agent
|
 |
|
The right system for every need
The MVA Systems (Metrohm VA Systems) are sets of instruments customized to meet particular demands. Choose the set that meets your requirements, connect it to a PC and a supply of nitrogen (if required for the particular system), and you are ready to go.
The systems provide different degrees of automation, depending on the number of samples to be analysed. They can be used as stand-alone or PC controlled systems.
|
| | |
AN-V-175 Antimony(III) in zinc plant electrolyte with chloranilic acid as complexing agent
|
|
The concentration of of Sb(III) in zinc plant electrolyte is determined by adsorptive stripping voltammetry (AdSV) with chloranilic acid as complexing agent. In this method high copper concentrations do not interfere. An approx. 10-fold excess of lead interferes, since it shows a signal close to the antimony. With the parameters given below the working range of this method is 1 - 30 µg/L antimony(III) with respect to the concentration in the measuring vessel.
|
|