Iron traces with 1-nitroso-2-naphthol

 MVA systems for VA1   The right system for every need
The MVA Systems (Metrohm VA Systems) are sets of instruments customized to meet particular demands. Choose the set that meets your requirements, connect it to a PC and a supply of nitrogen (if required for the particular system), and you are ready to go.

The systems provide different degrees of automation, depending on the number of samples to be analysed. They can be used as stand-alone or PC controlled systems.

 
AN-V-125 Iron traces with 1-nitroso-2-naphthol
The concentration of Fe is determined in water samples by adsorptive stripping voltammetry with 1-nitroso-2-naphthol as complexing agent. All reagents have to be added in the order as listed below. All reagents typically contain iron impurities. Therefore a subtraction of the reagent blank is recommended. Fe(II) and Fe(III) show different sensitivities. Therefore the sample should only contain one of the iron species. Ascorbic acid (Vitamin C) can be added to the measuring solution and to the Fe(III) standard solution if both Fe(II) and Fe(III) are present in the sample to determine the concentration of total iron. A final concentration of ascorbic acid of 0.002 mol/L is suitable.
 