Voltammetric determination of iron

 MVA systems for VA1   The right system for every need
The MVA Systems (Metrohm VA Systems) are sets of instruments customized to meet particular demands. Choose the set that meets your requirements, connect it to a PC and a supply of nitrogen (if required for the particular system), and you are ready to go.

The systems provide different degrees of automation, depending on the number of samples to be analysed. They can be used as stand-alone or PC controlled systems.

 
AB-317 Voltammetric determination of iron

This Application Bulletin describes two methods for the determination of iron at the Multi Mode Electrode.

Method 1, the polarographic determination at the DME, is recommended for concentrations of β(Fe) > 200 μg/L. For this method the linear range is up to β(Fe) = 800 μg/L.

For concentrations < 200 μg/L

Method 2, the voltammetric determination at the HMDE, is to be preferred. The detection limit for this method is β(Fe) = 2 μg/L, the limit of quantification is β(Fe) = 6 μg/L. The sensitivity of the method cannot be increased by deposition.

Iron(II) and iron(III) have the same sensitivity for both methods.

These methods have been elaborated for the determination of iron in water samples. For water samples with high calcium and magnesium concentrations such as, for example, seawater, a slightly modified electrolyte is used in order to prevent precipitation of the corresponding metal hydroxides. The methods can also be used for samples with organic loading (wastewater, beverages, biological fluids, pharmaceutical or crude oil products) after appropriate digestion.


 