Determination of aluminum in water samples by adsorptive voltammetry
|
 |
|
The right system for every need
The MVA Systems (Metrohm VA Systems) are sets of instruments customized to meet particular demands. Choose the set that meets your requirements, connect it to a PC and a supply of nitrogen (if required for the particular system), and you are ready to go.
The systems provide different degrees of automation, depending on the number of samples to be analysed. They can be used as stand-alone or PC controlled systems.
|
| | |
AB-186 Determination of aluminum in water samples by adsorptive voltammetry
|
This Bulletin describes the voltammetric determination of aluminum in water samples down to a concentration of 1 μg/L. An aluminum complex is formed with alizarin red S (DASA) and enriched at the HMDE. The following determination employs differential pulse adsorptive stripping voltammetry (DP-AdSV). Disturbing Zn ions are eliminated by addition of CaEDTA.
|
|