Home
Products
Applications
Service
Support
Company
Contact
 

 

Metrosep C 4 - 50/4.0

6.1050.450 Metrosep C 4 - 50/4.0
Cation separation column for the determination of monovalent and bivalent cations with improved peak forms, greater efficiency resulting in lower detection limits. PEEK column.
 
6.1050.450 Metrosep C 4 - 50/4.0

The Metrosep C 4 - 50/4.0 is the shortest separation column in the Metrosep C 4 product range. With a capacity of 6 µmol (K +), it is particularly suitable for very rapid separations. The low capacity makes it possible to quickly analyze the earth alkaline metals with their delayed elution. Thanks to the short retention times, applications that, in terms of analysis duration, were previously possible only with an FIA system (Flow Injection Analysis system) can now be transferred over to ion chromatography.

null
6.1050.450 Metrosep C 4 - 50/4.0
Column dimensions (mm) 4.0 x  50
Max. pressure 20 MPa
Maximum flow 2.0 mL/min
Standard flow 0.9 mL/min
Temperature range (°C) 20...60
Housing material PEEK
Carrier material Silica gel with carboxyl groups
Particle size (µm) 5
Column type 1
pH range 2...7
Storage In Standard eluent at 10 - 22°C
Weight 0.1
Regeneration

Disconnect the column outlet from the detector inlet.

The column must be rinsed with ultrapure water before and after the regeneration.

Organic contamination:
Rinse the column opposite the flow direction at a flow rate of 0.9 mL/min; 1 h with ultrapure water, 1 h with acetonitrile/water (40/60), 1 h with ultrapure water.

Inorganic contamination:
Rinse the column opposite the flow direction with 10 mM HNO 3 + 4 mM dipicolinic acid for 1 h at a flow rate of 0.9 mL/min.

Application range

Applications

  • Li + , Na + , K + , Rb + , Cs + , Mg 2+ , Ca 2+ , Sr 2+ , Ba 2+ , NH 4 +
  • Rapid separations
  • Simple sample matrices
Capacity

6 µmol (K+)

Organic modifier

0 - 100% (methanol not recommended)

Eluents

Nitric acid/dipicolinic acid eluent (standard eluent)

Nitric acid: 1.7 mmol/L; 3.4 mL/2 L (HNO 3, c = 1 mol/L)
Dipicolinic acid: 0.7 mmol/L; 234 mg/2 L

Nitric acid eluent

Nitric acid: 6.0 mmol/L; 12 mL/2 L (HNO 3, c = 1 mol/L)

6.1050.450 Metrosep C 4 - 50/4.0
Parts
6.2744.060 Threaded stopper
Options
6.1050.500 Metrosep C 4 Guard/4.0
6.1050.510 Metrosep C 4 S-Guard/4.0
6.1050.450 Metrosep C 4 - 50/4.0
6.1050.450 Metrosep C 4 - 50/4.0
8.107.5001

Brochure: Metrosep C 4

8.107.8016ML Leaflet for 6.1050.4x0 IC cation column Metrosep C 4/4.0, german/french/english/spanish