Metrosep C 4 - 250/4.0

6.1050.430 Metrosep C 4 - 250/4.0
Cation separation column for the determination of monovalent and bivalent cations with improved peak forms, greater efficiency resulting in lower detection limits. PEEK column.
 
6.1050.430 Metrosep C 4 - 250/4.0

The Metrosep C 4 - 250/4.0 is the cation column with the greatest capacity in the C 4 series. It is predestined for applications which require the highest separating efficiency. Samples with extreme differences in concentrations can be analyzed reliably with this column. The performance capability of the column is demonstrated, for example, in connection with the analysis of boiler feed water for which the requirement is the perfect quantification of 7 µg/L sodium in addition to 7 mg/L monoethanolamine (MEA). With the C 4 - 250/4.0, not only amines and transition metals but also alkaline and alkaline earth metals can be determined in a single run.

6.1050.430 Metrosep C 4 - 250/4.0
Column dimensions (mm) 4.0 x 250
Max. pressure 20 MPa
Maximum flow 2.0 mL/min
Standard flow 0.9 mL/min
Temperature range (°C) 20...60
Housing material PEEK
Carrier material Silica gel with carboxyl groups
Particle size (µm) 5
Column type 1
pH range 2...7
Storage In Standard eluent at 10 - 22°C
Weight 0.202
Regeneration

Disconnect the column outlet from the detector inlet.

The column must be rinsed with ultrapure water before and after the regeneration.

Organic contamination:
Rinse the column opposite the flow direction at a flow rate of 0.9 mL/min; 1 h with ultrapure water, 1 h with acetonitrile/water (40/60), 1 h with ultrapure water.

Inorganic contamination:
Rinse the column opposite the flow direction with 10 mM HNO 3 + 4 mM dipicolinic acid for 1 h at a flow rate of 0.9 mL/min.

Application range

Applications

  • Li +, Na + , K + , Rb + , Cs + , Mg 2+ , Ca 2+ , Sr 2+ , Ba 2+ , NH 4 + , Ni 2+ ,Zn 2+ , Co 2+ , Amine
  • NH 4 + , Monoethanolamine
  • Very good Na + /NH 4 + separation
  • NH 4 + , (CH 3)NH 3 + , (CH 3) 2NH 2 + , (CH 3) 3NH + , (CH 3) 4N + and the respective ethylammonium ions
  • Difficult separation problems
  • Great differences in concentration
  • Transition metals
Capacity

29 µmol (K+)

Organic modifier

0 - 100% (methanol not recommended)

Eluents

Nitric acid/dipicolinic acid eluent (standard eluent)

Nitric acid: 1.7 mmol/L; 3.4 mL/2 L (HNO 3, c = 1 mol/L)
Dipicolinic acid: 0.7 mmol/L; 234 mg/2 L

Nitric acid eluent

Nitric acid: 6.0 mmol/L; 12 mL/2 L (HNO 3, c = 1 mol/L)

6.1050.430 Metrosep C 4 - 250/4.0
Parts
2 pcs6.2744.060Threaded stopper
Options
6.1050.500 Metrosep C 4 Guard/4.0
6.1050.510 Metrosep C 4 S-Guard/4.0
6.1050.430 Metrosep C 4 - 250/4.0
8.000.6040EN

Improved cation separation thanks to a new column material (english)

AN-C-139

Cations and amines in the water-steam cycle

6.1050.430 Metrosep C 4 - 250/4.0
8.107.5001

Brochure: Metrosep C 4

8.107.8016ML Leaflet for 6.1050.4x0 IC cation column Metrosep C 4/4.0, german/french/english/spanish