iTrodes  60279300  

The iTrodes of the new intelligent electrode generation allow the storage of such important sensor data as

  • article and serial numbers,
  • calibration data,
  • calibration history,
  • working life and
  • calibration validity period.

All sensor data is read in automatically when the iTrode is connected to the Titrando. 

Mix-ups or editing errors are therefore eliminated. The electrode is identified automatically. If the type of electrode is not the same as that defined in the method the user is informed. This means that it is not possible to use the wrong electrode.

Monitoring functions allow the exclusion of electrodes whose calibration data lies outside the limits or whose calibration period has expired. The calibration data stored in the chip makes the electrode transferable; it does not need to be recalibrated each time that it is used with a different instrument.

For connection to your intelligent Titrator/Meter please use 854 iConnect.
 

 
6.0280.300 iEcotrode Plus

Excellent durability in routine work at a fair price.


  • Durable standard electrode for titration in various solutions
  • Fixed ground-joint diaphragm insensitive to contamination
  • For routine laboratory use
  • LL reference system with outstanding long-term stability
  • Quick and accurate signal response
  • Attractive price/performance ratio
 
6.0280.300 iEcotrode Plus
null
null
6.0280.300 iEcotrode Plus
Measuring unit pH
Indicator electrode type pH glass electrode
Shaft material Glass
Bridge electrolyte type c(KCl) = 3 mol/L
Diaphragm Fixed ground-joint diaphragm
Shaft diameter top (mm) 12
Resistance (kOhm) < 3 kOhm
Measuring range 0...13
Temperature range, short-term (°C) 0...80
Temperature sensor None
Reference electrolyte type c(KCl) = 3 mol/L
SGJ sleeve flexible SGJ sleeve
Membrane glass type special
Membrane resistance 50...300 MOhm
Temperature range, long-term (°C) 0...80
Plug for filling opening Orifice
Electrode slope < 0.97
Electrode zero point (mV) ± 15 mV
Shaft diameter bottom (mm) 12
Min. immersion depth (mm) 15
Internal reference electrode type Ag/AgCl cartridge
Indicator electrode shape Hemisphere
Weight 0.176